ETHICON ENDO-SURGERY, LLC. ENSEAL G2 STRAIGHT JAW 25CM; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number NSLG2S25
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # p9188a.Investigation summary: the device was received the upper jaw damaged at the proximal side.The device was connected to the generator and it was recognized.Because the jaw was damaged not all functional testing could be performed with the generator.The condition of the jaws prevented the functionality of the jaw.The device was unable to cycle open and close.A probable cause of the damage to the jaw could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event or Problem Description
The device was received with no complaint information.Attempts were made to reconcile the device with no success.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL G2 STRAIGHT JAW 25CM
Common Device Name
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7215709
Report Number3005075853-2018-07482
Device Sequence Number11020920
Product Code GEI
UDI-Device Identifier20705036014123
UDI-Public20705036014123
Combination Product (Y/N)N
PMA/510(K) Number
K112033
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2021
Device Catalogue NumberNSLG2S25
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2018
Initial Date Received by Manufacturer 01/04/2018
Initial Report FDA Received Date01/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-