Investigation conclusion: the customer's observation was not replicated in-house with retention products.Retention devices were tested with in-house drug-free donor urine and all devices showed expected negative results for all analytes including coc, mamp (met), thc, and mop analytes at the read time.No false positive results were observed during in-house testing.Manufacturing batch record review did not uncover any abnormalities.Based on the information available, there is no indication of a product deficiency and no corrective action is required.
|