MAUDE Adverse Event Report: MEDTRONIC MINIMED SENSOR MMT-7020B ENLITE3 US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED SENSOR MMT-7020B ENLITE3 US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-7020B
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 95 mg/dl and the sensor glucose was 40 mg/dl at the time of the incident.Customer¿s current blood glucose was 205 mg/dl and current sensor glucose value was 162 mg/dl.Customer reported that sensor glucose and blood glucose difference is not within acceptable range.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.Customer reported that insulin delivery was suspended due to sensor glucose values.Sensor glucose value that triggered the suspend event: 70 mg/dl.Threshold/low suspend limit in sensor settings: 70 mg/dl.The sensor will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR MMT-7020B ENLITE3 US
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key7215894
Report Number2032227-2018-00956
Device Sequence Number10798340
Product Code OZP
UDI-Device Identifier00643169704060
UDI-Public(01)00643169704060(017)20181220
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/20/2018
Device Model NumberMMT-7020B
Device Catalogue NumberMMT-7020B
Device Lot NumberL207
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2018
Initial Report FDA Received Date01/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age40 YR
Patient Weight178
-
-