The customer was reported via phone call that, the customer had high blood glucose of 538 mg/dl and customer treated it with a manual shot.Customer also reported that, the whole top of the reservoir compartment is missing and reservoir was not locking in place.Customer declined troubleshooting for the high blood glucose.Troubleshooting determined that, the insulin pump should be replaced.The customer advised to discontinue use of the pump and revert to a back-up plan.The insulin pump will be returned for analysis.
|
|
Insulin ump received with cracked case at the display window corner, broken reservoir tube lip, minor scratched lcd window, stained end cap sticker, stained address or serial number label, cracked belt clip slot, cracked case reservoir tube window corner, cracked lcd window, missing reservoir tube o-ring, cracked reservoir tube window and broken battery tube threads.Unable to perform the displacement, operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime or compromised force sensor system alarm and excessive no delivery test due to completely broken reservoir tube lip noted.
|