Per sus voluntary report mw5074228: it was reported that during an unspecified procedure, a fragment of the 0.018 inch diameter cope nitinol wire guide sheared off and was retained inside of the venous limb of the graft while exchanging to a bentson wire guide.The fragment was able to be successfully removed using alligator forceps and fluoroscopy guidance.Further information regarding patient outcome is unable to be requested, as customer contact information is unknown.
|
|
A review of the complaint history, device history record, documentation, drawing, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Numerous design verification and validation activities have been performed to ensure that this device meets design requirements.There is no information stating whether the pmg-18sp-60-cope-nt-st was withdrawn through a needle.The pmg-18sp-60-cope-nt-st is packaged in the 34cm(clear)slant-wire-guide-holder-w/caution-label (rm5082).Per dwg_80089-d-2, the 34cm(clear)slant-wire-guide-holder-w/caution-label includes the l_scor_rev.0 caution label.L_scor_rev.0 depicts a warning to not pull the distal spring coil portion of the wire guide through the needle tip.There is no information regarding other devices used in the procedure that may have contributed to the wire separation.There is no information regarding preparation of the device that may have contributed to the event.There is no information regarding resistance or human anatomical characteristics that may have contributed to the wire separation.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
|