• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. LINEAR HIP; STEM, HIP, FEMORAL, LINEAR LATERAL SIZE7

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. LINEAR HIP; STEM, HIP, FEMORAL, LINEAR LATERAL SIZE7 Back to Search Results
Model Number 400-32-000
Health Effect - Clinical Codes Bone Fracture(s) (1870); No Code Available (3191)
Date of Event 01/19/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Second revision surgery - due to the patient having a periprosthetic femur fracture.The surgeon removed and replaced the original stem and head.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to a periprosthetic femur fracture.The previous surgery and the revision detailed in this investigation occurred 3.8 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported components used in the previous surgery met design and manufacturing requirements.There was one (1) non-conforming material report (ncmr) associated with the part 400-32-000 (lot-891b1005) listed in the complaint.Non-conforming material report (ncmr) (b)(4) documents 2 parts for code v5 uniform surface finish (scratches on marking surface) disposition-rework 2 parts, acceptable after rework.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to a periprosthetic femur fracture.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to a periprosthetic femur fracture.The previous surgery and the revision detailed in this investigation occurred 3.8 years apart.The explanted stem was from the primary surgery and it was in-vivo for 16.2 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.The root cause of this complaint was a revision surgery due to a periprosthetic femur fracture.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR HIP
Common Device Name
STEM, HIP, FEMORAL, LINEAR LATERAL SIZE7
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7217032
Report Number1644408-2018-00075
Device Sequence Number120353
Product Code LPH
UDI-Device Identifier00888912076364
UDI-Public(01)00888912076364
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K974294
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2001
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/28/2007
Device Model Number400-32-000
Device Catalogue Number425-05-007
Device Lot Number733721
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/19/2018
Supplement Date Received by Manufacturer02/28/2018
03/07/2018
Initial Report FDA Received Date01/24/2018
Supplement Report FDA Received Date03/01/2018
03/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
425-05-007, LOT 733721
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
-
-