Model Number FUSION
Device Problems No Display/Image (1183); Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 01/03/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern. A medtronic representative went to the site to test the equipment. The representative was unable to replicate the reported issue. The hardware, software, and instruments passed the system checkout. The system was found to be fully functional. No parts were replaced, therefore no parts were returned for analysis.
 
Event Description
A medtronic representative reported that, while outside of a procedure, the navigation system screen lost display functionality. It was reported that the navigation system was restarted to resolve the issue. There was no patient present when this issue was identified. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFUSION NAVIGATION SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7217079
MDR Text Key98206009
Report Number1723170-2018-00375
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 01/24/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/21/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-