• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS DOT PROV SP 3F SL MAX; PERCUTANEOUS, IMPLANTED, LONG-TERM INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS DOT PROV SP 3F SL MAX; PERCUTANEOUS, IMPLANTED, LONG-TERM INTRAVASCULAR CATHETER Back to Search Results
Model Number N/A
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/29/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of rebs0167 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported by the facility that the patient had a picc placed on (b)(6) 2017.It was stated that approximately one month later, while under home care, the picc began to leak where the catheter inserted into the hub.There was no reported patient injury.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a break was inconclusive because no sample was returned to bas for evaluation.Based on the description of the reported event, possible contributing factors include sharp instrument damage, flexural fatigue, and over-pressurization; however, the lack of a returned sample prevented both confirmation of the reported event and identification of a root cause(s).Consequently, this complaint is inconclusive at this time.A lot history review (lhr) of rebs0167 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported by the facility that the patient had a picc placed on 11/30/2017.It was stated that approximately one month later, while under home care, the picc began to leak where the catheter inserted into the hub.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT PROV SP 3F SL MAX
Common Device Name
PERCUTANEOUS, IMPLANTED, LONG-TERM INTRAVASCULAR CATHETER
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7217250
Report Number3006260740-2018-00042
Device Sequence Number18859939
Product Code LJS
UDI-Device Identifier00801741134104
UDI-Public(01)00801741134104
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K162443
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Health Professional
Remedial Action Other,Notification
Type of Report Initial,Followup
Report Date (Section B) 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2017
Device Model NumberN/A
Device Catalogue NumberS1173108PD
Device Lot NumberREBS0167
Was Device Available for Evaluation? No
Event Location Home
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer12/28/2017
Initial Report FDA Received Date01/24/2018
Supplement Report FDA Received Date02/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age9 YR
-
-