• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: REVISION OPTICS RAINDROP NEAR VISION INLAY; CORNEAL INLAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

REVISION OPTICS RAINDROP NEAR VISION INLAY; CORNEAL INLAY Back to Search Results
Model Number 610-0001
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Corneal Clouding/Hazing (1878); Visual Impairment (2138)
Date of Event 11/28/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The inlay remains implanted in the patient's eye and is not available for evaluation.The device history record review of the manufacturing lot for this device was performed and there were no discrepancies or unusual findings related to the reported issue.Corneal haze and decreased vision are listed in the device labeling as known potential risks.(b)(4).
 
Event or Problem Description
The patient underwent implantation of the raindrop corneal inlay in the left eye on (b)(6) 2017.It should be noted that lasik surgery was performed on this eye concurrent with inlay implantation.On (b)(6) 2017, the patient presented with slight central and superior corneal haze in the operative eye and topical steroids were prescribed.The patient was examined on (b)(6) 2017 at which time his best corrected distance visual acuity (bcdva) decreased from 20/20 (preoperatively) to 20/70-; topical steroids were continued to treat the corneal haze.At last examination on (b)(6) 2018, the corneal haze resolved, bcdva improved to 20/25-2, and the inlay remained implanted.The surgeon believes that concurrent lasik may be a contributing factor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAINDROP NEAR VISION INLAY
Common Device Name
CORNEAL INLAY
Manufacturer (Section D)
REVISION OPTICS
25651 atlantic ocean dr.,
ste. a1
lake forest CA 92630 8835
MDR Report Key7217644
Report Number3005956347-2018-00016
Device Sequence Number123331
Product Code LQE
UDI-Device Identifier10850394006013
UDI-Public(01)10850394006013
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P150034
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date11/22/2019
Device Model Number610-0001
Device Catalogue NumberRD1-1
Device Lot Number003045
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/27/2017
Initial Report FDA Received Date01/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-