| Catalog Number |
121731056 |
| Medical Device Problem Code |
Disassembly (1168)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
01/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient heard a pop after a fall.Pinnacle liner disassociated from acetabular cup.Three (3) weeks post op.Products removed were 36x56 neutral altrx liner and 36+1.5 biolox ceramic head lot number is 8651122.There were no surgical delays.No additional concerns stated by surgeon.Patient consequence description:revision surgery for head and liner exchange.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #:(b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|