WRIGHTS LANE: SYNTHES USA PRODUCTS LLC 10MM TI CANNULATED TIBIAL NAIL-EX/375MM-STERILE; NAIL, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
04.004.455S |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Nonunion/Delayed-union Bone Fracture (2369); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Event date: it is unknown when the event occurred.Complainant part has not been returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a hardware removal and revision was performed on (b)(6) 2017 due to a non-union and broken plate.The patient was implanted with one (1) anterolateral distal tibia plate in the right tibia and one (1) 10mm cannulated tibia nail in the left tibia, and unknown number of screw on the both the left and right tibia in the month of (b)(6) 2017.The patient presented to the emergency room with pain in the leg on (b)(6) 2017.The surgeon brought him to surgery the following day for revision/repair non-unions.X-rays were taken and it was realized that the original plate was broken.Appeared to be non-infected hypertrophic non-unions.There were no inflammatory markers.All the hardware was successfully removed from the right tibia without any delay.The broken plate was removed and replaced with a non-locking medial plate as a medial strut, and placed a 3.5mm locking compression (lcp) anterolateral plate, with trinity and infuse.On the left the existing nail was removed and an exchange nail was implanted.The procedure was successfully completed with no surgical delay.No fragments were generated from the broken plate.The patient status was reported to fine.Concomitant device (s) reported: unknown locking screws (unknown part and lot number, unknown quantity).This complaint involves four (4) devices.This is report 2 of 4 for (b)(4).This report is for a tibial nail.
|
| |
|
Additional Manufacturer Narrative
|
|
Therapy date is (b)(6) 2017; exact date is unknown.A device history record (dhr) review was performed: manufacturing location: (b)(4), manufacturing date: 19-mar-2008, expiration date: 28-feb-2017: part #: 04.004.455s, lot#: 5733346 (sterile) ¿ 10mm ti cannulated tibial nail-ex/375mm- sterile.Quantity 6.: inspection sheet for in-process acceptance sheet, inspect dimensional, and inspection sheet for final inspection meet specification.Component parts reviewed: part 21012 bp80 lot: 5245281 reviewed.Raw material received from (b)(4).Certified test report received from (b)(4) for titanium meet specification.Raw material receiving/putaway checklist meets requirement.No non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Sterility documentation was reviewed and determined to be conforming.Corrected data: device implanted in (b)(6) 2017, exact date is unknown.Complainant device is not expected to be returned for manufacturer review/investigation.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Lot number, expiration date, and udi have been updated based on new lot number.Dhr review was completed.This device was manufactured and packaged sterile in (b)(4) under lot number 5733346.Manufacture date is 3/19/2008.Expiration date is 2/28/2017.Device was returned to synthes, unused, as part of recall z-2058-2013.Device was then forwarded to synthes (b)(4) where it was resterilized and repackaged.A new lot number was assigned, lot 7274593.Release to warehouse date for lot 7274593 is 3/22/2013.New expiration date is 2/28/2022.Manufacturing location: (b)(4) manufacturing date: 19-mar-2008 expiration date: 28-feb-2017 part #: 04.004.455s, lot#: 5733346 (sterile) ¿ 10mm ti cannulated tibial nail-ex/375mm- sterile.Quantity 6.Inspection sheet for in-process acceptance sheet, inspect dimensional 4004id431, and inspection sheet for final inspection meet specification.Component parts reviewed: part 21012 bp80 lot: 5245281 reviewed.Raw material received from (b)(4).Certified test report received from (b)(4) for (b)(4) meet specification.Raw material receiving/putaway checklist meets requirement.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.¿sterility documentation was reviewed and determined to be conforming.¿ dhr review for sterilization procedure only: part no.: 04.004.455s lot no.: 7274593 manufacturing location: (b)(4) supplier: (b)(4) release to warehouse date: 22.Mar.2013 expiry date: 28.Feb.2022 no ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Customer quality reviewed the x-rays provided.The 10mm ti cannulated tibial nail (04.004.455s, 5733346) was reported as being involved in an adverse event, but according to x-ray review, did not sustain any damage or contribute to the complaint condition.Based on x-ray review it is not possible to confirm a non-union.As a conservative measure, since it was reported that the revision surgery occurred due to a non-union, it will be confirmed, but replication of the complaint condition is inapplicable.Also, it was reported that the patient was a smoker and vitamin d deficient, both of which could have contributed to the complaint condition.Since the parts were not returned a drawing review and dimensional check could not be completed.Dhr review no ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.There is no allegation of device defect or deficiency associated with this complaint as reported for non-union.As there is no device malfunction alleged or identified, a dcrm review will not be performed due to the fact that there is no applicable hazard associated with this complaint to review within the dcrm.During the investigation no unidentified product design issues or discrepancies were observed that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|