Catalog Number NSEAL535RH
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/04/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # p9308t.Investigation summary: the device was received with the jaws stuck closed with a clip and dried tissue, between the jaws and the top of the i-blade fractured and slightly lifted.The device was connected to the generator and it was recognized.Because the i blade was damaged not all functional testing could be performed with the generator.The jaws were found attached to the instrument.In addition no pieces were found missing under microscope inspection.The condition of the blade prevented the functionality of the jaw.The jaw was unable to cycle open and close.A probable cause of the damage to the i blade could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.The batch history record was reviewed and no defects, nc¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event or Problem Description
It was reported that during a laparoscopic appendectomy, the handle would not close and open properly.The procedure was completed using an i-hook on an electrosurgery bovie unit.There were no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL 5MM ROUND TIP 35CM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7219019
Report Number3005075853-2018-07510
Device Sequence Number1773303
Product Code GEI
UDI-Device Identifier20705036000041
UDI-Public20705036000041
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K072177
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2019
Device Catalogue NumberNSEAL535RH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2018
Initial Date Received by Manufacturer 01/16/2018
Initial Report FDA Received Date01/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-