• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMPLANTECH ASSOCIATES, INC. GLUTEAL IMPLANT, ROUND, TEXTURED, SDM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPLANTECH ASSOCIATES, INC. GLUTEAL IMPLANT, ROUND, TEXTURED, SDM Back to Search Results
Model Number N/A
Medical Device Problem Code Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Method: the actual device was returned and gross visual examination and examination under magnification was performed.A review of manufacturing records and quality control records was also performed.Results: code for "none" selected as no more appropriate code was located.A tear in the circumference of the device was observed.Conclusion: breakage or damage to medical devices on implantation is a known risk of the procedure.Product labeling warns that a prosthesis that appears damaged should not be implanted.
 
Event or Problem Description
Complainant reported that they received a gluteal implant that contained bubbles.They proceeded to insert the implant into a keller funnel, and it ripped.The problem occurred before there was any patient contact, and they had a backup available so there was no impact to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLUTEAL IMPLANT, ROUND, TEXTURED, SDM
Common Device Name
GLUTEAL IMPLANT
Manufacturer (Section D)
IMPLANTECH ASSOCIATES, INC.
6025 nicolle street, suite b
ventura CA 93003
MDR Report Key7219175
Report Number2028924-2018-00001
Device Sequence Number18871530
Product Code MIB
UDI-Device IdentifierM724CCB6T841
UDI-PublicM724CCB6T841
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K982688
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/06/2022
Device Model NumberN/A
Device Catalogue NumberCCB6-T-84
Device Lot Number873720
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2018
Initial Report FDA Received Date01/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-