| Model Number |
CC4204A |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
| Date of Event |
12/12/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Collamer ultraviolet-absorbing posterior chamber single piece foldable intraocular lens (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a cc4204a collamer single piece lens, +19.0 diopter, in the patient's eye on (b)(6) 2017.The lens was explanted and exchanged on (b)(6) 2017, for a smaller diopter lens.The surgeon cut the lens to remove from the eye and there was no patient injury.The reporter indicated the lens was explanted due to a mispowered lens, it wasn't the patients desired outcome.The reporter indicated the lens was not mislabeled but was not sure if it was a possible miscalculation.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the lens was returned in liquid inside a specimen cup.Visual inspection found the lens torn in three pieces and the lens missing a piece of its haptic.(b)(4).
|
| |
|