BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) BLAZER® II XP; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number M0044500THK20
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned for analysis.The device was returned in a left curve position.There are bends located approximately 2cm and 5cm from the distal tip.Ring 2 and ring 3 seals are compromised.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.The left curve reached past the specified template area.The right curve test passed.X-ray confirmed two bends in the center support.The start of guide coil collapse was also noted in the handle.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is design specification as the device problem was traced back to the design specifications.(b)(4).
 
Event or Problem Description
Reportable based on device analysis completed on 12-jan-2018.It was reported that the catheter curve would not steer properly.A 7 / 110 / 2.5 / 8-8 blazer® ii xp was selected for use.During the procedure, it was noted that the catheter curve would not steer properly in one direction.The procedure was completed with another of the same device.No patient complications were reported.However, device analysis revealed that the seals of rings 2 and 3 were compromised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLAZER® II XP
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
MDR Report Key7219860
Report Number2134265-2018-00443
Device Sequence Number11066054
Product Code OAD
UDI-Device Identifier08714729452287
UDI-Public08714729452287
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P020025
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/11/2020
Device Model NumberM0044500THK20
Device Catalogue Number4500THK2
Device Lot Number0021123295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2018
Initial Report FDA Received Date01/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-