Model Number N/A
Device Problems Material Perforation (2205); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard. The device has not been returned to the manufacturer for evaluation, as the device was discarded after the event occurred. A lot history review (lhr) of rebt0432 showed three other similar product complaint(s) from this lot number. The complaints for this lot number (rebt0432) have been reported from the same facility.
 
Event Description
It was reported that the nurse used a powerglide catheter and stated that the guidewire was stiffer than normal and perforated and "kicked" the catheter. No patient hard was reported. This report addresses the third of four devices used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePOWERGLIDE MIDLINE CATHETER FULL KIT
Type of DeviceCATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas
MX
Manufacturer Contact
kelsey erickson
605 n. 5600 w.
salt lake city, UT 84116
8015225937
MDR Report Key7220682
MDR Text Key98533823
Report Number3006260740-2018-00055
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133856
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial
Report Date 01/25/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue NumberM118101
Device Lot NumberREBT0432
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location Hospital
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-