Model Number 03756230
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/13/2017
Type of Reportable Event Malfunction
Event or Problem Description
I was reported that when doctor used 2 polyaxial screws and 2 reduction long arm monoaxial screws.He inserted the screw on the left of s1.When he tried to use the cap inserter instrument to lock the cap into the screw, he found the cap is broken.There is a broken line on the cap.The doctor tried to get the broken cap.Out.However, because of the broken line, the cap was bigger than the head of the cap inserter, it was hard to get the cap out.Finally, after about 40 min,he used the cutter to cut the cap into pieces and got the pieces out, reinserted a new cap.
 
Additional Manufacturer Narrative
The blocker fractured during locking down the blocker.The universal tightener was used, which is just to be used for blocker insertion.For final tightening, the torque wrench must be used to ensure that excess torque is not applied to the blocker.The most likely root cause is excess force was applied to the blocker during tightening with either the driver not fully seated into the blocker or a deformed driver was used, causing the blocker to deform and eventually fracture.
 
Event or Problem Description
I was reported that when doctor used 2 polyaxial screws and 2 reduction long arm monoaxial screws.He inserted the screw on the left of s1.When he tried to use the cap inserter instrument to lock the cap into the screw, he found the cap is broken.There is a broken line on the cap.The doctor tried to get the broken cap.Out.However, because of the broken line, the cap was bigger than the head of the cap inserter, it was hard to get the cap out.Finally, after about 40min, he used the cutter to cut the cap into pieces and got the pieces out, reinserted a new cap.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA BLOCKER
Common Device Name
SPINAL INTERLAMINAL FIXATION ORTHOSIS.
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
-
allendale NJ 07401
MDR Report Key7221018
Report Number3005525032-2018-00005
Device Sequence Number16440504
Product Code KWP
UDI-Device Identifier04546540147257
UDI-Public(01)04546540147257
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K982494
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number03756230
Device Catalogue Number03756230
Device Lot NumberC4T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2018
Initial Date Received by Manufacturer 01/02/2018
Supplement Date Received by Manufacturer04/18/2018
Initial Report FDA Received Date01/26/2018
Supplement Report FDA Received Date05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
-
-