• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Hyperglycemia (1905); Vomiting (2144)
Date of Event 12/31/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The pump has not been returned to animas.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas alleging that the patient experienced a blood glucose of 502 mg/dl with vomiting associated with an alleged history/settings issue.Reportedly, it was unknown whether or not the patient continued with pump therapy or discontinued.The patient was treated by a health care provider at the hospital with intravenous fluids.During troubleshooting with customer technical support, it was noted that the pump was working as it should but the patient was not priming the pump.It was noted that basal delivery totals in the total daily dose did not match due to a cartridge change and the prime menu being accessed.All boluses were recorded as programmed.It was also noted that the patient bolused with delayed eating and there was a change in stress level.This complaint is being reported because the patient experienced hyperglycemia associated with user error and not a pump malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7221172
Report Number2531779-2018-01942
Device Sequence Number11357496
Product Code OYC
UDI-Device Identifier10840406100372
UDI-Public10840406100372
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other Caregivers
Type of Report Initial
Report Date (Section B) 01/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Device Age18 MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/02/2018
Initial Report FDA Received Date01/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age16 YR
Patient Weight85
-
-