| Model Number |
TRUETRACK |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results from results obtained of 165, 181, 145, 233 and 165 mg/dl.Customer is also concerned with erratic blood glucose test results from fasting back to back tests done during the call on (b)(6) 2018 of 126 and 151 mg/dl using truetrack meter.The customer's expected fasting blood glucose test result is 100 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the dining room.The test strip lot manufacturer's expiration date is 06/30/2019 and open vial date at time of call is two weeks ago.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results from results obtained of 165, 181, 145, 233 and 165 mg/dl.Customer is also concerned with erratic blood glucose test results from fasting back to back tests done during the call on (b)(6) 2018 of 126 and 151 mg/dl using truetrack meter.The customer's expected fasting blood glucose test result is 100 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the dining room.The test strip lot manufacturer's expiration date is 06/30/2019 and open vial date at time of call is two weeks ago.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.(b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Search Alerts/Recalls
|