| Model Number |
CSK-TC5 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Expiration date: na.
|
| |
|
Event or Problem Description
|
|
A report was received that the electrode fell out from its fixing head while taking off the guide.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint the electrode had a broken shaft was confirmed.Root cause of the failure is excessive external mechanical force placed upon the shaft which led to the complete separation of the shaft from the hub.
|
| |
|
Event or Problem Description
|
|
A report was received that the electrode fell out from its fixing head while taking off the guide.
|
| |
|