| Catalog Number |
471288 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer replaced aqua calibrator level 1 l/n 72853811 with l/n 72935498 to resolve the issue.
|
| |
|
Event or Problem Description
|
|
The customer stated erroneously low na (sodium) patient results were generated on their unicel dxc 600 pro synchron system using aqua calibrator level lot 72853811.The customer stated the patient sample results were reported outside of the laboratory, however the results were amended/corrected with no change to patient treatment.A review of the customer provided data confirms the instrument generated 1 (one) erroneously low na patient result.A review of the na quality control (qc) data provided shows recoveries for level 2 were greater than 2sd low of the instrument control mean on the event date.
|
| |
|