Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device would not power up.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The complainant was contacted for return of the device.The device has not been returned to zoll for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device.This supplemental medwatch report is also correcting information submitted on the initial medwatch report.Evaluation results: zoll medical corporation evaluated the device and the device performed to specification.The device was recertified and returned to the customer.Review of the device activity logs showed that the batteries were depleted but the device was still able to function and deliver therapy.No trend is associated with reports of this type.
|
| |
|