Catalog Number CLR222US |
Device Problem
Device Markings/Labelling Problem (2911)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint #: (b)(4).
Date sent to fda: (b)(4) 2018.
The actual sample was received for evaluation.
Evaluation of the received sample disclosed that the tyvek label showed an expiration date of (b)(4) 2012 , which is incorrect.
Per device history records review, the correct expiration date for subject product code (clr222us) and batch ((b)(4)) combination is (b)(4) 2019.
This is considered as incorrect and/or missing labeling condition.
|
|
Manufacturer Narrative
|
(b)(4).
To date, the device has not been returned.
If the product is returned for evaluation, any further information derived from the evaluation will be submitted in a supplemental 3500a form.
A review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
|
|
Event Description
|
It was reported that the expiration date on the blister pack of the topical skin adhesvie package is printed with an expiration date of 8/31/2012, whereas the expiration date on the outside of the box is 2019.
There was no patient involvement.
|
|
Search Alerts/Recalls
|