ABBOTT POINT OF CARE I-STAT CG4+ CARTRIDGE Back to Search Results
Catalog Number 03P85-50
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Overdose (1988); Respiratory Failure (2484)
Date of Event 01/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2018, abbott point of care was contacted by a customer regarding i-stat cg4+ cartridges that yielded suspected discrepant lactate result on a (b)(6) male patient with acute respiratory failure and overdose.There was no additional patient information available at the time of this report.Return product is not available for investigation.(b)(6).There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident # (b)(4).The investigation was completed on 02/06/2018.Retain product was tested and functioning according to specification.Return product was not available.
 
Event or Problem Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CG4+ CARTRIDGE
Common Device Name
CG4+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key7222801
Report Number2245578-2018-00023
Device Sequence Number11337848
Product Code KHP
UDI-Device Identifier10054749002266
UDI-Public10054749002266
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K982071
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2018
Device Catalogue Number03P85-50
Device Lot NumberD17274
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/09/2018
Supplement Date Received by Manufacturer02/06/2018
Initial Report FDA Received Date01/26/2018
Supplement Report FDA Received Date02/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
-
-