| Model Number |
ZCT450 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/27/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The intraocular lens remains implanted.Device evaluation: the lens remains implanted; therefore, it was not returned for investigation.The reported complaint was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The dfu provide the customer with proper usage instructions and guidelines.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that at the one-week post-operative visit on (b)(6) 2017, the doctor measured the intraocular lens (iol) axis in the right eye at 178 degrees.The iol was positioned at 8 degrees during surgery on (b)(6) 2017.There was no patient complication and/or related injury.The outcome doesn¿t significantly interfere with activities of daily life.No actions have been taken.The lens remains implanted.No additional information was provided to abbott medical optics.
|
| |
|