AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B1100-040
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual inspection was performed on the returned device.The reported balloon rupture and separation were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined that the reported difficulties were due to case circumstances.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a clot in the right femoral / right common iliac vein.Catheter assisted thrombolysis (ekos) was used.Post ekos, an armada 35 10 mm / 40 mm balloon was advanced from femoral to iliac vein.The armada was inflated twice and ruptured at 12 atmospheres near the previously implanted stents in the iliac vein.Angiography showed the distal end of the balloon was knotted and stuck on the guide wire and half the balloon was in the guiding catheter.The knotted portion of the balloon that was stuck on the guide wire was eventually safely removed when the sheath was upsized to a 14 french.Nothing was ever free-floating.The device was prepped according to the ifu.There was no resistance noted during advancement in the anatomy.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7223353
Report Number2024168-2018-00638
Device Sequence Number1023015
Product Code LIT
UDI-Device Identifier08717648154942
UDI-Public08717648154942
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 01/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2020
Device Catalogue NumberB1100-040
Device Lot Number70522G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2018
Initial Report FDA Received Date01/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age35 YR
Patient Weight69
-
-