(b)(4).Batch # n54h01.Device evaluation: the analysis showed that the 6tb45 device was received with no apparent damage and with a 6r45b cartridge loaded in the device.The reload was received partially fired 1/3 and with the reload lockout spring damaged.In addition, one (b - m52197) cartridge was received, the reload was fully fired and with the lockout spring normal.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.The returned device was tested for functionality with a test reload and it fired without any difficulties, the staple line was complete and the staples were noted to have the proper b-formed shape.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|