STRYKER SPINE-US RELIANCE T HANDLE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT. Back to Search Results
Model Number 48061000
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/02/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the quick release handles broke during surgery.All the broken fragments were retrieved.No adverse health consequences were reported.
 
Additional Manufacturer Narrative
Method: visual inspection, product history review, complaint history review, risk assessment; result: the reported event was confirmed via correspondence with the sales representative and device inspection of the returned product.The inner prongs of the instrument were confirmed to be fractured upon visual inspection.No manufacturing related issues were found upon product history review.According to correspondence with the sales representative, the instrument has been used hundreds of times.Based on the information provided and the age of the device (8 years), the possible root cause of the reported event is due to normal wear and tear.
 
Event or Problem Description
It was reported that the quick release handles broke during surgery.All the broken fragments were retrieved.No adverse health consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANCE T HANDLE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT.
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key7224298
Report Number0009617544-2018-00023
Device Sequence Number16437793
Product Code LXH
UDI-Device Identifier07613327263671
UDI-Public(01)07613327263671
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number48061000
Device Catalogue Number48061000
Device Lot Number108615
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2018
Initial Date Received by Manufacturer 01/02/2018
Supplement Date Received by Manufacturer04/24/2018
Initial Report FDA Received Date01/29/2018
Supplement Report FDA Received Date05/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-