COOK ENDOSCOPY ACROBAT® 2 CALIBRATED TIP WIRE GUIDE; OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number G47616
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: our evaluation of the product said to be involved confirmed the report.There is wire guide coating damage near the distal end.The wire guide is bent at approximately 6 cm with a kink at the distal tip, and bent again at approximately 15.5 cm.Approximately 2.4 cm to 4.1 cm from the distal end, the wire guide covering has folded over itself.Approximately 4.1 cm to 5.8 cm from the distal end is a section of bare core wire.Due to the condition of the returned device it cannot be determined if any sections of the coating are missing.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.The instructions for use instruct the user to do the following: "prior to removing wire guide from holder, flush with 30 cc of sterile water." failure to flush the wire guide can result in damage to the wire guide.The instructions for use instruct the user to do the following: "flush endoscope accessory channel and/or lumen of device with sterile water, then insert wire guide floppy end first.Note: for best results, wire guide should be kept wet, if applicable." failure to flush the endoscope channel can result in damage to the wire guide.Prior to distribution, all acrobat 2 calibrated tip wire guides are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook acrobat 2 calibrated tip wire guide.The tip of the wire stripped about 2 cm back from the tip when trying to gain access into the proximal common bile duct.The user did not notice [the coating damage], until the wire was pulled out.The user then used a different wire to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACROBAT® 2 CALIBRATED TIP WIRE GUIDE
Common Device Name
OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key7226094
Report Number1037905-2018-00032
Device Sequence Number896072
Product Code OCY
UDI-Device Identifier00827002476165
UDI-Public(01)00827002476165(17)200317(10)W3838365
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG47616
Device Catalogue NumberAWG2-35-450
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date01/05/2018
Device Age10 MO
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2018
Initial Report FDA Received Date01/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-