• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRATHANE COPE NEPHROURETEROSTOMY SET; LJE CATHETER, NEPHROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRATHANE COPE NEPHROURETEROSTOMY SET; LJE CATHETER, NEPHROSTOMY Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
The customer reported that the catheter portion of the ultrathane cope nephroureterectomy set was leaking from around the hub of the device.The customer confirmed there were no patient adverse events which resulted form the product problem, and no additional procedures were necessitated by the issue.The procedure being performed was a nephroureteral stent placement, and the leakage reportedly occurred during the initial placement of the device.The device is reportedly available for return; however, as of the date of this report, no device has yet been received for evaluation.
 
Additional Manufacturer Narrative
Investigation evaluation: a review of the complaint history, dimensional verification, functional test, device history record, documentation, and visual inspection of the returned device was conducted during the investigation.The device was returned and leak testing has confirmed the reported failure.Overall, all measurements matched specifications.A document based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.Based on the information provided and the results of our investigation, a definitive root cause cannot be determined at this time, however appropriate measures have been initiated to address this failure mode.Per the [quality engineering] risk assessment, no further action is required.Monitoring will continue to be performed for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information or changes at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE COPE NEPHROURETEROSTOMY SET
Common Device Name
LJE CATHETER, NEPHROSTOMY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7226752
Report Number1820334-2018-00160
Device Sequence Number1016059
Product Code LJE
UDI-Device Identifier00827002481763
UDI-Public(01)00827002481763(17)200920(10)8231107
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberULT8.5-8.5-24-NUCL-B-RH
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 01/12/2018
Supplement Date Received by Manufacturer06/01/2018
Initial Report FDA Received Date01/29/2018
Supplement Report FDA Received Date06/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-