| Model Number |
U3575510 |
| Medical Device Problem Codes |
Break (1069); Kinked (1339)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation for an angioplasty procedure, an alleged kink and break was visible at the distal part of the catheter shaft.Another balloon was used to perform the procedure.There was no patient contact.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation for an angioplasty procedure, an alleged kink and break was visible at the distal part of the catheter shaft.Another balloon was used to perform the procedure.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the device history records were reviewed.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.A visual inspection found a kink on the catheter shaft.Therefore, the investigation is confirmed for the reported kink.However, the investigation is unconfirmed for the reported break, as no breaks were identified anywhere along the length of the device.It is unknown whether handling techniques by the user as they removed the catheter from the packaging or prepped the device may have contributed to the kinked catheter.It is unlikely that the kink is manufacturing related, as a 100% visual inspection for catheter kinks is performed per spec.Based upon the available information a definitive root cause has not been determined.Labeling review: the current ifu (instructions for use) states: precautions: carefully inspect the catheter prior to use to verify that catheter has not been damaged during shipment and that its size, shape, and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.
|
| |
|