Model Number 1516-40-906
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Edema (1820); Pain (1994); Loss of Range of Motion (2032); Swelling (2091); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Date of Event 10/30/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Legal medical records received jan 8, 2018.Right total knee replacement from (b)(6) 2014 (a revision of a competitor unicompartmental arthroplasty) was in turn revised (b)(6) 2017 for pain, recurrent effusions, and poor range of motion of 10-80 degrees.Revising surgeon identified that the tibial baseplate component was internally malrotated approximately 10-15 degrees.There was also noted premature polyethylene wear of the far posterior medial aspect of the tibial insert.Surgeon also indicated that the patella component was undersized, with about 12 mm of lateral native patella bone uncovered and making contact with the femur component.Femur, tibial baseplate and patella implant components were all well-fixed.All products were revised.Doi: (b)(6) 2014.Dor: (b)(6) 2017, right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: udi: (b)(4).Added: patient identifier, date of event, description of event or problem, other relevant history and evaluation codes (patient) no code available was used to capture medical device removal and surgical intervention.Corrected: explant date and evaluation codes (patient).
 
Event or Problem Description
Corrected code for swelling and decreased joint range of motion.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy as not synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary = no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  added: b7, d11, h6 (no code available (3191) is used to capture the device revision or replacement).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FB INSRT SZ 9 6MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key7227494
Report Number1818910-2018-52369
Device Sequence Number997971
Product Code JWH
UDI-Device Identifier10603295050490
UDI-Public10603295050490
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2015
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 01/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2018
Device Model Number1516-40-906
Device Catalogue Number151640906
Device Lot Number367192
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/08/2018
Supplement Date Received by Manufacturer11/28/2018
12/09/2019
03/10/2020
05/20/2020
Initial Report FDA Received Date01/30/2018
Supplement Report FDA Received Date11/28/2018
12/20/2019
03/23/2020
05/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 7 CEM; ATTUNE MEDIAL DOME PAT 35MM; ATTUNE PS FEM RT SZ 9 CEM; SMARTSET GHV GENTAMICIN 40G; SMARTSET GHV GENTAMICIN 40G; SMARTSET GHV GENTAMICIN 40G
Outcome Attributed to Adverse Event Required Intervention;
Patient Age42 YR
-
-