Model Number 60726807
Health Effect - Clinical Codes Air Embolism (1697); Patient Problem/Medical Problem (2688)
Date of Event 01/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Bayer service performed a check of the injector and found it to perform to specifications.The subject disposables used during the injection were discarded at the site and the site was unable to provide the lot numbers.The site declined clinical applications assistance.This information does not constitute an admission that the device, the company or its employees caused or contributed to this reportable event.
 
Event or Problem Description
The customer reported the following: a (b)(6) year old female with a history of a periumbilical abscess underwent a ct of the abdomen and pelvis with contrast while connected to a stellant injector (serial number (b)(4)).Post injection, the radiologist visualized an indeterminate amount of air in the right ventricle on the images.No unusual (technical) issues were reported during the procedure to explain the presence of air.The patient was asymptomatic.The patient was kept for 24 hour observation and discharged to home the next day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLANT DUAL CT INJECTOR
Common Device Name
CT INJECTION SYSTEM
Manufacturer (Section D)
BAYER MEDICAL CARE, INC.
one bayer drive
indianola PA 15051
MDR Report Key7227628
Report Number2520313-2018-00019
Device Sequence Number19138214
Product Code DXT
UDI-Device Identifier00616258008943
UDI-Public(01)00616258008943(11)170825
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K082905
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 01/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number60726807
Device Catalogue NumberSCT D W/ CERTEGRA
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/11/2018
Initial Report FDA Received Date01/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age27 YR
-
-