| Model Number |
7207607 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the screw was found disintegrated when it was trying to screw it in.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Examination was not possible, as the device has not been returned.The investigation could not draw any conclusions about the reported event without the return of the device.A review of the ncmr database was performed which confirmed no abnormalities were reported with this lot during manufacture.A complaint history review identified no additional complaints for this manufactured lot.
|
| |
|
Additional Manufacturer Narrative
|
|
One 7x25mm biorci reverse thread screw was returned for evaluation.Visual assessment of the screw confirmed the reported complaint of breakage.Approximately 5mm of the distal threads have broken off.The broken portion was returned in three pieces of varying sizes.Dimensional assessment of the screws major diameters and wall thickness was performed and found to be within print specifications.Due to the limited clinical details provided we are unable to provide a root cause for this failure.This investigation could not identify any evidence of product contribution to the reported complaint.
|
| |
|