| Model Number |
H7493808020300 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/15/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluated by mfr.: the device was not returned for analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in a coronary artery.A 3.0mm x 20mm quantum¿ maverick¿ balloon catheter was advanced for post-dilatation.The balloon was initially inflated at 12 atmospheres; however, during the second inflation at 18 atmospheres for 20 seconds, the balloon ruptured.The device was completely removed from the patient and the procedure was completed with a non-bsc balloon catheter.No patient complications nor injuries were reported.
|
| |
|