| Model Number |
AB62181002 |
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Information (3190)
|
| Date of Event |
12/22/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information of conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.However, there are no other complaints on this lot number.Checking the quality databases revealed no anomaly in connection with the described event.Investigation which didn't reveal any anomaly recorded during production.
|
| |
|
Event or Problem Description
|
|
According to the available information, the patient tore off his catheter when he will stood up.The catheter is broken inside urethra.Clinical consequence : the patient is transferred to the operating room.And other catheter was been in place.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow up mdr is created to document the additional patient information and event information received, and the conclusion of the investigation.After receiving this complaint, we searched for other complaint and we didn't find other complaint on this lot.Checking the quality databases revealed no anomaly in connection with the described defect.Unfortunately no sample is available from the customer and we cannot go further than the documentary investigation which didn't reveal any anomaly recorded during production.According to the ifu : warning/precautions : this type of device must only be used by trained and experienced professionals.Any use other than stated indications is under the responsibility of the physician follow-up : regular monitoring should be implemented to ensure that no side effect occurs, that the catheter is working properly and in particular that it is draining properly.Monitoring should also check that the balloon is still inflated and the catheter correctly placed.No other vigilance has been found on the item ab6218 since ce marking of this product.Our clinical assessment concluded : prostatic catheters must be inserted by a healthcare professional and post-operative nursing should be implemented.It is recommended that patients be educated to understand clearly the catheterization to cooperate properly and adapt their behavior.If not possible, close monitoring of the patient should be implemented.The isolated incident of rupture of x-flow prostatic catheter was possibly caused by insufficient fixation of device in a geriatric patient, maybe agitated, who tore of his catheter while standing up as mentioned by the complainant.Based on the above this complaint is not confirmed.
|
| |
|
Event or Problem Description
|
|
According to the available information, the patient tore off his catheter when he stood up.The catheter is broken inside urethra.Clinical consequence : the patient is transferred to the operating room.And other catheter was been in place.Additional information received indicated the catheter was broken 1/3 from the balloon side.The patient was (b)(6) years old and suffers from fecal impaction at the bladder eye.There is no clinical consequences before the transfer to surgery, then the patient had a local anaesthesia for the surgery.The balloon was inflated in the bladder.The surgery was a bladder fibroscopy, with removal of the end of the catheter remained in place.Patient was catheterized with another catheter without difficulty.
|
| |
|