| Model Number |
1011-0139-03 |
| Medical Device Problem Code |
Failure to Discharge (1169)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
12/30/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a (b)(6) male patient, the associated defibrillator was unable to discharge using these internal handles.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty pin.The handles were scrapped and the customer received a replacement.It's important to mention these handles are over a 11 years of age and are recommended to be replaced after either 100 sterrad or 50 autoclave cycles of sterilization.Zoll also recommends a functional inspection prior to each use.Analysis for reports of this type has not identified an increase in trend.
|
| |
|