DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE II INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.110 |
Device Problem
Device Stops Intermittently (1599)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/11/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.
Reliability engineering evaluated the device.
A visual and functional assessment was performed on the device which found that it worked intermittently.
During repair, it was determined that the electronic control unit (ecu) was damaged and had no voltage output.
Therefore, the reported condition was confirmed.
The assignable root cause was determined to be due to normal use over time.
If information is obtained that was not available, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that during service and repair pre-testing, it was discovered that the small battery drive device worked intermittently.
The event was not related to surgery.
There was no patient involvement.
There were no injuries or medical intervention associated with this event.
All available information has been disclosed.
If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|