MEDTRONIC NAVIGATION, INC. (LOUISVILLE) PATIENT TRACKER 9734887XOM NON-INVASIVE; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9734887XOM
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During the onsite inspection the medtronic representative found that the wire on the patient tracker was tied in a knot and damaged with the wires exposed.The rep replaced the part.The replaced part has not been received by the manufacturer for further analysis.A successful system checkout was performed which verified that the issue had been resolved.
 
Event or Problem Description
A medtronic representative reported that during a catheter placement procedure the site registered the patient, but when they moved toward the navigate screen the patient tracker was not visible in the tracking view, but a stylet was visible when they brought it in the field of view.The case continued without navigation.There was no patient impact reported and the delay in surgery was less than an hour.Surgery proceeded as planned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT TRACKER 9734887XOM NON-INVASIVE
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
MDR Report Key7228631
Report Number1723170-2018-00459
Device Sequence Number16436485
Product Code HAW
UDI-Device Identifier00643169488120
UDI-Public00643169488120
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K141833
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 01/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number9734887XOM
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2018
Initial Report FDA Received Date01/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Patient Weight78
-
-