• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GREATBATCH MEDICAL MOBICATH BIDIRECTIONAL GUIDING SHEATH, SMALL CURVE; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GREATBATCH MEDICAL MOBICATH BIDIRECTIONAL GUIDING SHEATH, SMALL CURVE; INTRODUCER, CATHETER Back to Search Results
Model Number 1000182-001
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/05/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the catheter was leaking from the hub shortly after the mobi-cath was inserted.The catheter was replaced and the issue resolved.
 
Additional Manufacturer Narrative
The sheath device was returned for evaluation.Visual analysis of the sheath did not reveal any discrepancies, nor were any discrepancies noted upon actuation of the deflection mechanism.Pressure testing did not reveal any leaks from the valve.No root cause could be determined.Device analysis and review of manufacturing records did not reveal any discrepancies related to this device failure mode.No corrective action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBICATH BIDIRECTIONAL GUIDING SHEATH, SMALL CURVE
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
GREATBATCH MEDICAL
2300 berkshire lane
minneapolis MN 55441
MDR Report Key7228917
Report Number2183787-2018-00004
Device Sequence Number1222505
Product Code DYB
UDI-Device Identifier00821329400995
UDI-Public00821329400995
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K130559
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date (Section B) 01/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/10/2018
Device Model Number1000182-001
Device Catalogue NumberD140010
Device Lot NumberVN3880868
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/08/2018
Supplement Date Received by Manufacturer01/08/2018
Initial Report FDA Received Date01/30/2018
Supplement Report FDA Received Date04/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-