NUVASIVE, INC. NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 15355500
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product has been returned for evaluation nor were radiographs provided to confirm the reported event.Patient was reported to be doing well post revision procedure.Potential adverse events and complications: ".As with any major surgical procedures, there are risks involved in orthopedic surgery." ".Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s); nonunion or delayed union." ".These devices can break when subjected to the increased load associated with delayed union or nonunion.Internal fixation appliances are load-sharing devices that hold bony structures in alignment until healing occurs.If healing is delayed, or does not occur, the implant may eventually loosen, bend, or break.Loads on the device produced by load bearing and by the patient's activity level will dictate the longevity of the implant." ".Patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." post-operative warnings: ".During the postoperative phase, it is of particular importance that the physician keeps the patient well informed of all procedures and treatments.Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration as well as to other complications." no product returned for evaluation.
 
Event or Problem Description
On (b)(6) 2017, a patient underwent posterior spinal fusion at t4-iliac levels.On (b)(6) 2018, a fracture of one of the rods was detected at l5-s1 level.A revision procedure was performed on (b)(6) 2018, where both rods at l5-s1 level were cut and screws at s1-iliac levels were removed.Patient was reported to be doing well post revision procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk blvd
san diego CA 92121
MDR Report Key7229205
Report Number2031966-2018-00010
Device Sequence Number19138215
Product Code NKB
UDI-Device Identifier00887517510143
UDI-Public00887517510143
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K143684
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number15355500
Device Lot NumberCT62636
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2018
Initial Report FDA Received Date01/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-