ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT300
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zxt300 22.0 diopter intraocular lens (iol) was implanted in the patient's operative eye on (b)(6) 2017.It was later explanted on (b)(6) 2017, because of a myopic issue and the visual outcome was not achieved.There was no incision enlargement, vitrectomy, or sutures used.The replacement lens was the same model, but different diopter of 20.0.Reportedly, there was no patient injury and the patient is doing fine post-operatively.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 1/23/2018.Device returned to manufacturer? yes.Device evaluation: the sample was returned to the manufacturer.Visual inspection at 12x microscope magnification was performed and it can be seen that the product is damaged and incomplete, most probably to make explant possible.The complaint cannot be confirmed.Manufacturing record review: manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed no similar complaints were received for this production order number to date.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the intraocular lenses.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7229270
Report Number9614546-2018-00091
Device Sequence Number16451244
Product Code POE
UDI-Device Identifier05050474606425
UDI-Public(01)05050474606425(17)220420
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date04/20/2022
Device Model NumberZXT300
Device Catalogue NumberZXT300U220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2018
Initial Date Received by Manufacturer 01/03/2018
Supplement Date Received by Manufacturer03/12/2018
Initial Report FDA Received Date01/30/2018
Supplement Report FDA Received Date04/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-