Model Number ZXT150
Medical Device Problem Code Invalid Sensing (2293)
Health Effect - Clinical Codes Visual Disturbances (2140); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Explant date: if explanted, give date: not applicable as there is no indication of a lens explant.(b)(4).Attempts have been made to obtain missing information; however, to date no response has been received.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zxt150 22.5 diopter intraocular lens (iol) was implanted into the right eye (od) of the patient on (b)(6) 2017.It was reported that the patient can not see near and or computer vision, and the patient feels that their vision is worse than before surgery.To date, lens remains implanted.No further information was provided.
 
Additional Manufacturer Narrative
(b)(4).Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key7229774
Report Number9614546-2018-00096
Device Sequence Number19091651
Product Code POE
UDI-Device Identifier05050474605251
UDI-Public(01)05050474605251(17)211103
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date11/03/2021
Device Model NumberZXT150
Device Catalogue NumberZXT150U225
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/03/2018
Supplement Date Received by Manufacturer06/04/2018
Initial Report FDA Received Date01/30/2018
Supplement Report FDA Received Date06/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-