Analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.A lot history review revealed 9 additional reocclusion complaints were associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual samples were not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was not related to the study device or procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.Furthermore, thrombosis and pain are listed as potential adverse events, per the ifu.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
It was reported through a clinical registry that during the index procedure, two lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the right mid and distal superficial femoral artery (sfa) and right popliteal artery.Approximately eight days after the index procedure, the patient reportedly experienced symptoms of numbness and pain, and the patient was hospitalized, in which it was found that the patient's right distal sfa and popliteal vessels had re-occluded with thrombosis.A thrombolysis followed by a balloon angioplasty a week later were performed and the hcp deemed they were successful.The investigator assessed that the event was not related to the study device or procedure.The samples were discarded by the user facility and are not available for evaluation.No adverse patient effects were reported.This is one of two products involved with the reported event and the associated manufacturer¿s report number is 3006513822-2018-00018.
|