• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ALTRX +4 10D 28IDX44OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ALTRX +4 10D 28IDX44OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 122128144
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Not Applicable (3189)
Date of Event 01/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Der states that the patient had a left total hip done in (b)(6) in (b)(6) 2013.This was with a corail stem and pinnacle cup.Developed pain a year ago and was brought in for revision.The 44 mm liner was disassociated with the cup.It looked like a +4 10 degree liner but the lip of the liner was in pieces and no numbers could be seen.The 28 mm +1.5 ceramic head showed wear on the head.A new 44 mm +4 10 defeee liner was implanted with a ts ceramic 28 mm +5 head.Doi: (b)(6) 2013; dor: (b)(6) 2018; left hip.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRX +4 10D 28IDX44OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582
MDR Report Key7230299
Report Number1818910-2018-52417
Device Sequence Number1249742
Product Code LPH
UDI-Device Identifier10603295015987
UDI-Public10603295015987
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K062148
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 01/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number122128144
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/11/2018
Supplement Date Received by Manufacturer02/22/2018
Initial Report FDA Received Date01/31/2018
Supplement Report FDA Received Date02/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
-
-