| Catalog Number |
122128144 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Not Applicable (3189)
|
| Date of Event |
01/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Der states that the patient had a left total hip done in (b)(6) in (b)(6) 2013.This was with a corail stem and pinnacle cup.Developed pain a year ago and was brought in for revision.The 44 mm liner was disassociated with the cup.It looked like a +4 10 degree liner but the lip of the liner was in pieces and no numbers could be seen.The 28 mm +1.5 ceramic head showed wear on the head.A new 44 mm +4 10 defeee liner was implanted with a ts ceramic 28 mm +5 head.Doi: (b)(6) 2013; dor: (b)(6) 2018; left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation summary: no device associated with this report was received for examination.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|