| Catalog Number |
RONYX50012UX |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
12/01/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
During the procedure a resolute onyx rx drug eluting stent was deployed in the rca, at 12 atm.It was reported that the stent embolized into the guide catheter.The guide was retracted however, the stent remained in the right subclavian.The stent was snared and pulled back into the guide catheter and removed from the body intact.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|