| Catalog Number |
254500978 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint: (b)(4).Investigation summary: examination of the returned device confirms the reported event of a damaged balseals.Hhe (health hazard evaluation)/qrb (quality review board) (b)(4) recommended a device correction which was initiated on june 12, 2015.Capa-(b)(4) determined the likely root cause to be related to misuse, and requires mandatory sales training on proper use and application of the device by depuy sales consultants.Capa-(b)(4) will also monitor the mandatory field training.The need for further corrective action was not indicated.Continue to monitor per post market surveillance (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that size 8 spring broken on articulating surface and size 7 nub broke off near spring articulating surface.No surgical delay.No pieces of broken instrumentation left in patient.
|
| |
|
Additional Manufacturer Narrative
|
|
There is no new information to report at this time.Per fda request, this follow-up is being submitted to fill the gap in follow-up sequence numbers.
|
| |
|
Event or Problem Description
|
|
It was reported that size 8 spring broken on articulating surface and size 7 nub broke off near spring articulating surface.No surgical delay.No pieces of broken instrumentation left in patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: this product was reported in error.No patient death, serious injury or evidence of reportable product malfunction occurred.Depuy considers the investigation closed at this time.Should any additional information be received to change the outcome of the performed investigation, the complaint will be re-opened.
|
| |
|