BAYER MEDICAL CARE, INC. STELLANT CT SYRINGE KIT; ANGIOGRAPHIC SYRINGE KIT Back to Search Results
Model Number 60728818
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis received and examined the returned tube set.Visual examination identified the particulate as degraded resin.Degraded resin can collect on the extrusion tooling and be introduced during the manufacturing process.A 12 month review of complaint history determined the frequency to be (b)(4) complaints per million (cpm) for similar defects.
 
Event or Problem Description
The customer reported the following: after opening the stellant ct disposable kit the customer observed a black particulate inside the tubing.The customer elected not to use the syringe kit.No injury or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLANT CT SYRINGE KIT
Common Device Name
ANGIOGRAPHIC SYRINGE KIT
Manufacturer (Section D)
BAYER MEDICAL CARE, INC.
1 bayer drive
indianola PA 15051
MDR Report Key7230619
Report Number2520313-2018-00018
Device Sequence Number1994783
Product Code DXT
UDI-Device Identifier00616258005218
UDI-Public(01)00616258005218(11)171208(17)211208(10)8538707
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K082905
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/08/2021
Device Model Number60728818
Device Catalogue NumberSSS-CTP-SPK
Device Lot Number8538707
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2018
Initial Report FDA Received Date01/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-