Model Number 980
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The service engineer (se) inspected the device and verified the reported issue.The se replaced the graphical user interface (gui) and the breath delivery (bd) central processing unit (cpu).The se updated the software to the current revision and upgraded the battery firmware.The se performed calibrations and extended self-testing on the device and all tests passed.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that, while in use on a patient, a 980 ventilator alarmed with error codes, was inoperable and began backup ventilation.The patient was removed from the ventilator and placed on an alternate ventilator.The patient was not harmed or injured as a result of the reported event.
 
Additional Manufacturer Narrative
Device evaluation: the graphical user interface (gui) printed circuit board (pcb) and the breath delivery (bd) pcb were returned for failure investigation.The components were visually inspected and functionally tested.The reported failure could not be replicated.There was no fault found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key7230814
Report Number8020893-2018-00053
Device Sequence Number13725758
Product Code CBK
UDI-Device Identifier10884521171633
UDI-Public10884521171633
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K162738
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number980
Device Catalogue Number980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2018
Initial Date Received by Manufacturer 01/16/2018
Supplement Date Received by Manufacturer03/13/2018
Initial Report FDA Received Date01/31/2018
Supplement Report FDA Received Date03/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-