• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASPEN SURGICAL PRODUCTS, CALEDONIA BARD-PARKER CARBON RIB-BACK BLADES SIZE 11, STERILE; BARD-PARKER BLADE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASPEN SURGICAL PRODUCTS, CALEDONIA BARD-PARKER CARBON RIB-BACK BLADES SIZE 11, STERILE; BARD-PARKER BLADE Back to Search Results
Model Number 371111-150
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 12/01/2017
Event Type  Injury  
Manufacturer Narrative
The following controls are in-place to mitigate ¿broken blade¿ condition at aspen surgical (b)(4) site: heat treatment in-process at the beginning and end of each lot are inspected for dimension integrity using a pin gauge, flatness gauge and perforation length gauge, ductility test, and hardness test.Heat treatment quality inspections at the beginning and end of each lot are inspected for dimension integrity using a pin gauge, flatness gauge and perforation length gauge, ductility test, and hardness test.The 100% visual inspection by certified personnel is performed to segregate nonconforming material, including broken blades, prior to assembly and packaging process.Each associate performing 100% inspection process is certified and re-certified on an annual basis on blade defect criteria¿s.Capa (b)(4) was initiated to evaluate current controls in place corrective actions identified were completed and capa closed on 10/26/2015.Capa (b)(4) was also initiated to continue in the monitoring and improvement of current control.Device not available.
 
Event Description
During an arthroscopy procedure, item number 371111 broke during the surgical incision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD-PARKER CARBON RIB-BACK BLADES SIZE 11, STERILE
Type of Device
BARD-PARKER BLADE
Manufacturer (Section D)
ASPEN SURGICAL PRODUCTS, CALEDONIA
6945 southbelt dr. s.e.
caledonia MI 49316
Manufacturer (Section G)
SAME AS ABOVE
Manufacturer Contact
jordan hackert
6945 southbelt dr. s.e.
caledonia, MI 49316
MDR Report Key7231140
MDR Text Key98679871
Report Number1836161-2018-00007
Device Sequence Number1
Product Code GES
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number371111-150
Device Lot Number0127941
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age59 YR
Patient Weight116
-
-